Independent analysis puts a WWB semaglutide lot at 95.3%, against 94.2% on the certificate
The report states the gradient, the wavelength and the integration threshold, which is more than most.
TheCompound Journal
Reporting on incretins, compounding & the peptide supply chain
Testing Dispatch
The supplier confirmed the determination and supplied the method behind its own figure inside a week.
Two laboratories analysing vials from the same dulaglutide lot have returned figures 3.7 percentage points apart. Medutest reported 96.5%; Janoshik reported a lower number on a shallower gradient with a tighter integration threshold. Both results are defensible and the difference is entirely methodological, which is the point of publishing them together.
The Journal purchased 15 vials from the same lot at retail and had four of them analysed separately, in order to distinguish vial-to-vial variation from method variation. Vial-to-vial agreement was within 0.4 percentage points. Method-to-method disagreement was an order of magnitude larger.
The Journal purchased 15 vials from the same lot at retail and had four of them analysed separately, in order to distinguish vial-to-vial variation from method variation. Vial-to-vial agreement was within 0.4 percentage points. Method-to-method disagreement was an order of magnitude larger.
Purity became this trade’s single figure of merit because it is the cheapest measurement to obtain and the easiest to compare. Peptide content, which determines how much drug is actually in a vial, costs several times as much and is quoted as a priced addition by most of the twenty companies in the Journal’s dossier programme rather than carried as a standard release test. Sterility and endotoxin are separate disciplines again, and are neither tested nor claimed.
Dr Emiko Tanabe, mass spectrometrist, said the gap was predictable from the methods alone. "You can reconstruct almost the whole difference from the integration threshold. What you cannot reconstruct is why anyone thinks a number without a threshold is comparable to a number with one."
CPC’s dossier has been updated with this determination and with the company’s response to it. The dossier’s ordering box records the lead times, the minimum quantities and what a buyer should ask for at the point of order.
The report states the gradient, the wavelength and the integration threshold, which is more than most.
The Journal submitted the sample and paid for the analysis. The vendor was told in advance.
The Journal submitted the sample and paid for the analysis. The vendor was told in advance.
Documentation practice is the only part of vendor quality a buyer can assess before purchase.
Where two methods disagree, the conservative convention is to report the lower figure. It is not universal, and whether a laboratory follows it belongs on the report.
A reminder that a purity figure is the output of a method, and that methods differ.